I. Introduction:
DP-LDZ4-2 automatic balance centrifuge has beautiful appearance, large capacity, small size and complete functions. Its performance is stable, the speed is adjustable and it can automatically adjust the balance, low temperature rise, high efficiency and wide applicability. It is suitable for qualitative analysis of serum, plasma and urea in pharmaceutical products, blood stations, clinical trials and biochemical laboratories.
Second, the automatic balance centrifuge technical parameters:
model
Product name
Rotating speed
capacity
Maximum relative departure
power supply
DP-LDZ4-2
Automatic balance
Centrifuge
0-4000rpm
20ml×8
2350xg
220V 50HZ
Third, before use:
1. Read this manual carefully before using the automatic balance centrifuge.
2. The working environment temperature of the automatic balance centrifuge is 5-40 °C, the relative humidity should not exceed 80%, and there is no conductive dust, explosive gas and corrosive gas.
3, the workbench should be level and stable to prevent vibration, the work room should be clean and tidy, dry and well ventilated.
4. Wipe the inner cavity and rotor before cleaning.
5. Before work, put the empty centrifuge tube evenly and let the machine run at the highest speed for 1-2 minutes. No abnormality can be used for normal work.
Fourth, the automatic balance centrifuge instructions:
1. Prepare for the “pre-use instructionsâ€.
2. Place the tube with the same weight and fit into the centrifuge hole and close the centrifuge lid.
3. Turn on the power, turn on the power switch, and the power indicator lights.
4. Adjust the speed knob to set the desired speed, ranging from 600 rpm to 4000 rpm.
5. Adjust the timer to set the centrifugation time: 5-120 minutes.
6. After the work is finished, turn off the power and clean the whole machine.